Tektronix DPO2000 - MSO2000 Oscilloscopes – Mixed signal oscilloscope – Factory adjustment procedure – How to calibrate the DC voltage and sine wave signal – Main board parts description – Exploded view – test and measurement equipment repair and service
Category: Test and Measurement Equipment Repair and Service Contents of this article Factory adjustment procedure DC and sine wave signal calibration Mainboard parts and expoled view Tektronix DPO 2000 - MSO 2000 Oscilloscopes Factory Adjustment Procedure To perform the factory adjustment procedure, complete these steps: 1. Connect the oscilloscope to an AC power source. NOTE: You must connect the oscilloscope and the test equipment to the same AC power circuit. Connect the oscilloscope and test instruments to a common power strip if you are unsure of the AC power circuit distribution. Connecting the oscilloscope and test instruments to separate AC power circuits can result in offset voltages between the equipment, which can invalidate the adjustment procedure. 2. Power on the oscilloscope and warm it up for 20 minutes. 3. Power on the oscilloscope calibrator and warm it up for 20 minutes. 4. Connect the active heads from the calibrator to ...